欢迎进入一品仪器设备有限公司网站!
一品仪器 - 产品展示
yipin instrument-Products
4新闻中心
您的位置:首页  ->  新闻中心

Test Results


​ 

                        Test Results

​Table 1 Table of measured temperature values for the lowest and highest temperature detection points in each space

Temperature setting: 85°C Indication: 84.9°C~85.2°C      Temperature setting: 150°C Indication: 149.9 °C to 150.2 °C   

Detection point location

Measured value

Upper deviation

Lower deviation

 

Detection point location

Measured value

Upper deviation

Lower deviation

upper layer

A

84.5

0.0

-0.5

 

upper layer

A

149.5

0.0

-0.5

B

84.6

0.0

-0.4

B

149.5

0.0

-0.5

C

84.3

0.0

-0.7

C

150.3

+0.3

0.0

D

85.1

+0.1

0.0

D

149.6

0.0

-0.4

Middle layer

O

85.5

+0.5

0.0

 

Middle layer

O

149.2

0.0

-0.8

Lower layer

E

84.4

0.0

-0.6

 

Lower layer

E

149.3

0.0

-0.7

F

84.8

0.0

-0.2

F

149.6

0.0

-0.4

G

85.3

+0.3

0.0

G

148.6

0.0

-1.4

H

84.1

0.0

-0.9

H

148.9

0.0

-1.1

The test point concludes:

Temperature uniformity: ±2°C O-point temperature fluctuation: ±0.3 °C

 

 

The test point concludes:

Temperature uniformity: ±2 °C O point temperature fluctuation: ±0.3°C

 

Table 2 Actual temperature measurement table of the lowest and highest temperature detection points in each space

Left:Temperature setting: -60°C Indication: 59.5°C~-60.5°C

Right:Temperature setting: 0°C Indication: 0.6 °C to -0.9 °C    

Detection point location

Measured value

Upper deviation

Lower deviation

 

Detection point location

Measured value

Upper deviation

Lower deviation

upper  layer

A

-59.8

0.0

-0.2

 

upper layer

A

0.6

0.2

-0.4

B

-59.6

0.0

-0.4

B

0.6

0.4

-0.2

C

-59.5

0.0

-0.5

C

0.9

0.8

-0.1

D

-59.5

0.0

-0.5

D

0.7

0.4

-0.3

Middle layer

O

-60.1

+0.1

0.0

 

Middle layer

O

0.8

0.6

-0.2

Lower layer

E

-60.5

0.0

-0.5

 

Lower layer

E

0.6

0.2

-0.4

F

-59.5

0.0

-0.5

F

0.8

0.7

-0.2

G

-60.1

+0.1

0.0

G

0.9

0.8

-0.1

H

-59.6

0.0

-0.4

H

0.8

0.6

-0.2

The test point concludes:

Temperature uniformity: ±2°C O-point temperature fluctuation: ±0.3 °C

 

 

The test point concludes:

Temperature uniformity: ±2 °C O point temperature fluctuation: ±0.3°C

 

 

 

 

[返回]   
Copyright © 2017 东莞市一品仪器设备有限公司 版权所有 本站素材及相关资源均来源互联网,如有侵权请速告知,我们将会在24小时内删除*【Gmap】 访问量:后台管理 粤ICP备16117078号
服务热线
0769-84393814
产品符合ISO、ASTM、DIN、EN、GB、BS、JIS、ANSI、UL、等国际测试标准
专注试验设备多年
关闭
点击这里给我发消息
在线咨询
点击这里给我发消息
技术咨询